| 3474
ELECTROMAGNET SYSTEM - HIGH PRECISION |
APPLICATIONS:
When integrated with a high stability Magnet Power
Supply and a high precision Teslameter, the Model
3474 can be applied to FMR, EPR, and NMR studies,
VSM magnetic property measurements, magnetic standards,
and sensor calibration. A typical system configuration
follows. Other options include Vertical Mounts,
Motorized Rotating Drives, and NMR Teslameters
If a simple system with field reversal is required,
please go the the
System with Field Reversal page.
Please download our Minimum
Facilities Requirements paper for inportant
information regarding system installiation, site
set-up and materials required. |
|
| Specifications |
| Pole
gap |
Variable
from 0 to 160mm. |
| Pole
Face Diameter |
150mm
(250mm, 200mm 100mm, 75mm, 50mm,
and 25mm pole caps are also available). |
| Operating
Field |
±1.9T
at a pole gap of 51mm (2.0 inch) with ±140A excitation. |
| Field
Uniformity |
±150ppm
over a 10mm diameter disc in the median
plane. |
| Current
Stability |
±1ppm
over 30 Minutes, ±3ppm over 8 hours. |
| Current
Control Resolution |
18
bit. |
| Field
Readout Resolution |
20
bit (1µT up to 3.0T field). |
| Computer
Control |
IEEE-488
interface with LabView. |
| Cooling |
Water,
26l/min (7 US gpm - for Magnet and
Power Supply). |
| AC
Power |
208Vac,
three phase, 48/62Hz, 56A/phase |
| AC
Power (Equipment Rack) |
115Vac,
single phase, 50/60Hz, 15A |
| |
| System
Components |
| 1A |
1 |
3474-140 |
H-frame
Electromagnet, 250mm, 140A
coils |
26,440 |
| 1B |
1 |
17801713 |
Pole
pair, 150mm face diameter |
1,300 |
| 1C |
1 |
17801730 |
45°
Mounting Angle Pair |
540 |
| 1D |
1 |
11801800 |
Rolling
Base |
2,230 |
| 1E |
1 |
16900000 |
Cable,
Current and Interlock, 160A, 5m |
610 |
| 1F |
1 |
853-160A/80V |
Power
Supply, 160A, 80V, with current
reversal and IEEE-488 interface, ±10ppm
over 8 hours |
38,300 |
| Electromagnet
sub system: |
$69,490 |
| |
| 2A |
1 |
DTM-151-DG |
Digital
Teslameter, IEEE-488 Interface,
20-bit |
4,390 |
| 2B |
1 |
MPT-141-7s |
Hall
Probe, 4 range, 0.3T, 0.6T, 1.2T,
3.0T |
3,080 |
| 2C |
1 |
11901230 |
Probe
Holder, aluminum, transverse |
150 |
| 2D |
1 |
11901250 |
Probe
Mount, 4-way adjustable |
730 |
| Teslameter
sub system: |
$8,350 |
| |
| 3A |
1 |
RC-351930 |
Bench
height 19 inch rack |
2,600 |
| 3B |
1 |
17000025 |
Rack
Mounting Panel, 1 Teslameter |
110 |
| 3C |
1 |
17000058 |
Panel
Mount instrument Support Bracket |
70 |
| 3D |
1 |
CIF24-1m |
Cable,
IEEE-488, 1m |
160 |
| 3E |
1 |
CIF24-4m |
Cable,
IEEE-488, 4m |
200 |
| Rack
sub system: |
$3,140 |
| |
| 4A |
1 |
RMCS-P4 |
Rackmount
Computer System, Pentium 4.
| |
3U Rackmount Computer with rails, 2GB RAM, 250GB Hard Drive, DVD±R, Windows XP Pro, Keyboard & Optical Mouse, 17" LCD Monitor and 25' extension cables for keyboard, mouse & monitor. |
|
$2,240 |
| 4A |
1 |
778032-01 |
PCI-GPIB Interface for Windows 2000/XP/Vista |
800 |
| 4B |
1 |
LVMCTRL |
LabView
Magnet Control software |
1,070 |
| 4C |
1 |
776671-09 |
LabView for Windows, base package, 1Yr service |
1,800 |
| Computer
control sub system: |
$5,910 |
| |
| |
| 5A |
1 |
MC500-A1-E1-
H2-2J2(¾)-B¾ |
Water
Chiller, Water Cooled, Refrigerated
Re-circulating. Includes water temperature
and
pressure controls, dual outlets, filter, flow
meter
and low flow interlock. |
$18,200 |
| |
| 5A |
1 |
Installation |
Installation
& Training, up to 3 days at customers site
|
$6,970 |
| |
| Complete
Electromagnet System: |
$112,060 |
|
Specifications and pricing are subject to change