- Main Unit - 3045
|
| Electrical
|
| Range |
0.08T
to 7.0T (3.4MHz to 300MHz proton NMR) |
| Resolution |
0.01ppm |
| Accuracy |
Better
than 5ppm |
| Repeatibility |
Better
than 0.1ppm |
|
Frequency
Stability
Over Time
Temperature
Aging
|
Better than 0.05ppm over 3 days
± 0.1ppm from -25°C to +70°C
± 1ppb / day |
Maximum
Number
of Probes |
32
(special order to 64 or 96) |
| Readings |
All
probes simultaneously |
| Read
Cycle Time |
5
seconds, typical |
| Operating
Temperature |
10°C
to 40°C |
| Storage
Temperature |
-
20°C to + 70°C |
| Environmental
Field |
<
0.5T |
| Computer
Interface |
RS-232-C |
| Power
Input |
??V,
??A, 12W |
| Mechanical
|
| Enclosure |
5mm
aluminum plate |
| Dimensions |
270
x 120 x 75mm |
| Weight |
3kg |
| Connectors |
| External
Power Input |
Lemo
8-pole, type EGG.2B.308.CLL |
| Probe
Array |
DB25,
female, with positive locking mechanism |
| RF
to Probe Array |
Lemo
2-pole, type EGG.2B.302.CLL |
| RS-232-C |
DB25,
female |
| Remote |
Lemo
5-pole, type EGG.0B.305.CLL |
|
|
|
| |
|
Specifications
- Power Supply - 3046
|
| Electrical
|
| Power
Input |
85
to 240VAC, 40VA |
| Power
Output |
??V,
??A, ??W |
| Fuse |
2
amp, slow blow |
| Operating
Temperature |
10°C
to 40°C |
| Storage
Temperature |
-
20°C to + 70°C |
| Operating
Field |
<
0.05T |
| Controls |
| Reset
Switch |
Resets
power supply and main unit |
| Mechanical
|
| Enclosure |
Extruded
aluminum and plastic |
| Dimensions |
210
x 120 x 50mm |
| Weight |
1.5kg |
| Connectors |
| Power
Input |
IEC
standard input socket with fuse holder |
| Power
Output |
Lemo
8-pole, type EGG.2B.308.CLL |
|
|
|
| |
|
Specifications
- Probe Array - 3048
|
| Electrical
|
| Center
Frequency |
User
defined |
| Frequency
Range |
±
2%, typical, from the Center Frequecny |
Tunable
Frequency  |
3.4
to 300MHz, tuned at factory to the Center
Frequency |
Maximum
Number of
Probes |
32
(special order to 64 or 96) |
| Probe
to Probe Offset |
Better
than ± 0.2ppm after normalization |
| Probe
Position Accuracy |
Better
than ± 0.3mm
|
| Operating
Temperature |
10°C
to 40°C |
| Storage
Temperature |
-
20°C to + 70°C |
| Mechanical
|
| Enclosure |
|
| Geometry |
As
defined by application requirements |
| Dimensions |
As
defined by application requirements |
| Weight |
1.5kg |
| Cable
Length |
3
meters, standard |
|
|
|
 |
|