Specifications
- THM 7025
|
Electrical
|
 Range |
19.99mT,
199.9mT, or 1999mT, manual or auto |
 Resolution |
0.01mT,
0.1mT, or 1mT, depending on range |
 Accuracy |
±
2% of reading or ± 3 counts |
 Temp
coefficient of gain |
±
0.05%/°C |
 Display |
3½
digit LCD |
 Update
rate |
0.4
seconds or hold |
 Measurement
mode |
Single
or three axis |
 3
axis calculation |
 |
 Operating
temperature |
0°C
to +40°C |
 Operating
field (meter) |
0.1T
(or 1.5T with battery removed) |
 Computer
interface |
RS-232,
9600 baud, 8 bits, no parity, 1 stop,
no XON/XOFF |
 External
power input |
9Vdc,
400mA |
 Battery |
Type
SLM9V, 9V, lithium |
   Controls |
   On/Off |
Switches
on or off power |
   Range |
Manually
changes range |
   Bz |
Toggle
from single to three axis mode |
   Zero |
Zero
probe offset |
   Hold |
Hold
reading while button is depressed |
Mechanical
|
 Enclosure |
High
strength ABS plastic |
 Dimensions,
meter |
160
x 80 x 30 mm |
 Dimensions,
probe |
12
x 12 x 100 mm |
 Probe
cable length |
1.5
meters standard |
 RS-232
cable length |
3
meters standard, 30 meters max |
 Weight |
250g |
 Storage
temperature |
-
20°C to + 60°C |
 Connectors |
   External
power input |
3.5mm
mini co-axial socket |
   Sensor |
8
position DIN socket |
   RS-232 |
3.5mm
mini stereo jack |
|
|
System
components (standard)
|
 THM
7025 |
Hand
held three axis teslameter |
 THS
7025-10 |
Probe
with 1.5 meter cable |
 SLM9V |
9
volt lithium battery |
 HPS
7025-10 |
AC
power adaptor (specify 110V or 220Vac) |
 SC
7025-13 |
RS-232C
cable, 3 meters |
 ZFC
7025-16 |
Zero
field chamber |
 CC
7025 |
Carrying
case |
 UM
7025 |
Users
manual |
|
|
The below products are mechanically and functionally similar |
| Narda Model |
Metrolab Model |
| ETM-1 |
THM-702550 |
Last updated on:
31 October, 2007
|