| Specifications
- YM12-3-2-0.2T & YM12-3-2-2T |
|
Electrical
Characteristics |
| Model
Number |
YM12-3-5-0.2T |
YM12-3-5-2T |
YM12-3-5-5T |

Full-Scale Field (Max) |

±0.2T (±0.4T) |

±2.0T (±4.0T) |
±5.0T |
| Differential
Voltage Outputs: (Bx, By, & Bz) |
±10.0V |
| Sensitivity |
50V/T
(5mV/G) |
5V/T
(0.5mV/G) |
2V/T
(0.2mV/G) |
| Accuracy |
±0.1%
(B=0.1T) |
±0.1%
(B=1T) |
±0.1%
(B=1T) |
| Non-Linearity |
<1%
(B=0.2T) |
<1%
(B=2T) |
<2%
(B=5T) |
| Long
Term Stability |
<1%/10
years |
| Temperature
Coefficient of Sensitivity |
<100ppm/°C |
| Offset
(B=0T) |
±10mV
(±2G) |
±1mV
(±2G) |
±0.5mV
(±2.5G) |
| Temperature
Coefficient of Offset (B=0T) |
<1mV/°C
(0.2G) |
<0.1mV/°C
(0.2G) |
<0.1mV/°C
(0.2G) |
| Output
Noise and Ripple (Peak, 0.01 to 100Hz) |
<±0.5mV
(±0.1G) |
<±0.06mV
(±0.12G) |
<±0.05mV
(±0.1G) |
| Output
Resistance (Short Circuit Protected) |
<10W |
Frequency
Response
1% Error
-3dB |
>1kHz
5kHz |
Temperature
Operation (within spec.)
Storage |
10 to 50°C
-20 to 85°C |
| Power |
±12V,
50mA |
| Mechanical |
| Electronics
(Module E) |
Connectors
Power
Signal Output |
DIN SFV50, 5 Pole
DIN KFV70, 7 Pole |
| Dimensions
(mm) |
95
x 120 x 37 |
Weight
Module E
Signal Output Cable
|
470g 43g/m |
| Material |
Aluminum |
| Probe
(Module H) |
| Dimensions
(mm) |
5.0
x 2.0 x 14.0mm |
Weight
H-Module (field sensor)
H-Module to E-Module Cable |
2g
42g/m |
| Material |
Ceramic
with Epoxy Cap |
| Field
Sensitive Volume (Dx,
Dy, Dz) |
0.125
x 0.10 x 0.125mm |
| Center
of the Field Sensitive Volume |
2.5
x 1.45 x -1.0mm |
Angular
Accuracy of the Axis
To the Reference Ceramic |
±0.5° |
Cable
Length
Standard
Maximum |
5 meters
10 meters |
| Electromagnetic
Testing |
RF
Conducted Disturbances: IEC/EN 61000-4-6, ENV 50141
Radiated Electromagnetic Field: ENV
50140
Pulse Modulated Electromagnetic Field:
ENV 50140
Electrical Fast Transient Burst: IEC/EN
61000-4-4
Electrostatic Discharge: IEC/EN 61000-4-2*
Test Conditions:
| |
The transducer was tested using Senis 'S12'
power supply, which conforms to the directives
73/23/EEC and 89/336/EEC.
The output cable used was Senis 'CO15' shielded
output cable with ferrite RI-14-28 (25MHz,
177W). The 'GND'
output contact was grounded.
For the ESD test the flexible cable section
(thin wires) and the probe head were sheltered
inside the plastic protection tube delevered
with the transducer. |
|
| Drawings: |
|
|