This Electromagnet System provides a bipolar field with computer control. The Electromagnet has small size and low mass and can be readily mounted with any field direction. It is appropriate for laboratory or production-line measurements of Hall effect, magneto-optic and other magnetic characteristics. Options are available for axial holes for light entry and exit.
If a unipolar system is required, please go the the Unipolar System page.
Specifications
Pole gap
Variable from 0 to 75mm.
Pole Face Diameter
20mm or 40mm (reversible poles)
Operating Field
±0.65T at a pole gap of 20mm with ±6A true bipolar excitation. (Higher fields with smaller gaps.) Intermittant operation to 8A excitation for approximately 0.8T with a 20mm pole gap.
Field Uniformity
±0.5% over an 18mm diameter disc in the median plane.
Current Stability
0.02% over 8 hours.
Current Control Resolution
12 bit.
Field Readout Resolution
12 bit (0.1mT up to 0.6T field).
Computer Control
IEEE-488 interface with LabVIEW.
Cooling
Water, 1l/min (0.26 US gpm).
Air cooling up to approx. 3.5A excitation.
AC Power
115V/230Vac selectable, single phase, 50/60Hz, 9.5A, (@ 115V).
Specifications are subject to change without notice.