This Electromagnet System achieves field reversal with computer control. It is particularly suitable for Hall effect studies, magneto-optical measurements, and high temperature superconductor investigations. Options are available for higher fields at reduced pole gaps (e.g. 1.7T at a 25mm gap with a 50mm pole face diameter), axial holes for light entry and exit, and unipolar power supplies. For larger pole gap requirements up to 100mm refer to the 3472-50 which has 50A rather than 70A coils.
If a true bipolar system is required, please go the the True Bipolar System page.
Specifications
Pole gap
Variable from 0 to 82mm.
Pole Face Diameter
100mm (75mm, 50mm, and 25mm pole caps are
also available).
Operating Field
±0.9T at a pole gap of 51mm (2.0 inch) with ±70A excitation. Intermittant operation to 100A excitation for approximately 1.1T.
Field Uniformity
±0.5% over a 30mm diameter disc in the median plane.
Current Stability
0.05% over 8 hours.
Current Control Resolution
14 bit.
Field Readout Resolution
12 bit (0.2mT up to 1.2T field).
Computer Control
IEEE-488 interface with LabVIEW.
Cooling
Water, 6l/min (1.6 US gpm).
Air cooling up to approx. 20A excitation.
AC Power (Power Supply)
187-242V, 21A, 47-63Hz ac, 3 phase
AC Power (Equipment Rack)
115Vac, single phase, 50/60Hz, 15A
Specifications are subject to change without notice.
GMW can provide a wide range of Electromagnet Systems for laboratory and industrial applications.